Abstract

Attenuation correction (AC) is essential for quantitative positron emission tomography (PET) images. Attenuation coefficient maps (μ-maps) are usually generated from computed tomography (CT) images when PET-CT combined systems are used. If CT has been performed prior to PET imaging, pre-acquired CT can be used for brain PET AC, because the human head is almost rigid. This pre-acquired CT-based AC approach is suitable for stand-alone brain-dedicated PET, such as VRAIN (ATOX Co. Ltd., Tokyo, Japan). However, the headrest of PET is different from the headrest in pre-acquired CT images, which may degrade the PET image quality. In this study, we prepared three different types of μ-maps: (1) based on the pre-acquired CT, where namely the headrest is different from the PET system (μ-map-diffHr); (2) manually removing the headrest from the pre-acquired CT (μ-map-noHr); and (3) artificially replacing the headrest region with the headrest of the PET system (μ-map-sameHr). Phantom images by VRAIN using each μ-map were investigated for uniformity, noise, and quantitative accuracy. Consequently, only the uniformity of the images using μ-map-diffHr was out of the acceptance criteria. We then proposed an automated method for removing the headrest from pre-acquired CT images. In comparisons of standardized uptake values in nine major brain regions from the <superscript>18</superscript>F-fluoro-2-deoxy-D-glucose-PET of 10 healthy volunteers, no significant differences were found between the μ-map-noHr and the μ-map-sameHr. In conclusion, pre-acquired CT-based AC with automated headrest removal is useful for brain-dedicated PET such as VRAIN. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
18650333
Volume :
16
Issue :
4
Database :
Complementary Index
Journal :
Radiological Physics & Technology
Publication Type :
Academic Journal
Accession number :
173804799
Full Text :
https://doi.org/10.1007/s12194-023-00744-z