Back to Search Start Over

Surgical navigation for guidewire placement from intraoperative fluoroscopy in orthopaedic surgery.

Authors :
Mekki, L
Sheth, N M
Vijayan, R C
Rohleder, M
Sisniega, A
Kleinszig, G
Vogt, S
Kunze, H
Osgood, G M
Siewerdsen, J H
Uneri, A
Source :
Physics in Medicine & Biology; 11/7/2023, Vol. 68 Issue 21, p1-13, 13p
Publication Year :
2023

Abstract

Objective. Surgical guidewires are commonly used in placing fixation implants to stabilize fractures. Accurate positioning of these instruments is challenged by difficulties in 3D reckoning from 2D fluoroscopy. This work aims to enhance the accuracy and reduce exposure times by providing 3D navigation for guidewire placement from as little as two fluoroscopic images. Approach. Our approach combines machine learning-based segmentation with the geometric model of the imager to determine the 3D poses of guidewires. Instrument tips are encoded as individual keypoints, and the segmentation masks are processed to estimate the trajectory. Correspondence between detections in multiple views is established using the pre-calibrated system geometry, and the corresponding features are backprojected to obtain the 3D pose. Guidewire 3D directions were computed using both an analytical and an optimization-based method. The complete approach was evaluated in cadaveric specimens with respect to potential confounding effects from the imaging geometry and radiographic scene clutter due to other instruments. Main results. The detection network identified the guidewire tips within 2.2 mm and guidewire directions within 1.1°, in 2D detector coordinates. Feature correspondence rejected false detections, particularly in images with other instruments, to achieve 83% precision and 90% recall. Estimating the 3D direction via numerical optimization showed added robustness to guidewires aligned with the gantry rotation plane. Guidewire tips and directions were localized in 3D world coordinates with a median accuracy of 1.8 mm and 2.7°, respectively. Significance. The paper reports a new method for automatic 2D detection and 3D localization of guidewires from pairs of fluoroscopic images. Localized guidewires can be virtually overlaid on the patient's pre-operative 3D scan during the intervention. Accurate pose determination for multiple guidewires from two images offers to reduce radiation dose by minimizing the need for repeated imaging and provides quantitative feedback prior to implant placement. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
00319155
Volume :
68
Issue :
21
Database :
Complementary Index
Journal :
Physics in Medicine & Biology
Publication Type :
Academic Journal
Accession number :
173095623
Full Text :
https://doi.org/10.1088/1361-6560/acfec4