Back to Search Start Over

A combined autofluorescence and diffuse reflectance spectroscopy for mucosa tissue diagnosis: Dual‐distance system and data‐driven decision.

Authors :
Liu, Dongyuan
Zheng, Jie
Zhang, Qi
Zhang, Limin
Gao, Feng
Source :
Journal of Biophotonics; Oct2023, Vol. 16 Issue 10, p1-9, 9p
Publication Year :
2023

Abstract

Combined autofluorescence (AF) and diffuse reflectance (DR) spectroscopies have been expected to offer enhanced diagnostic accuracies for noninvasive early detection of mucosa lesions, that is, oral cavity carcinoma and cervical carcinoma. This work reports on a hybrid AF and DR spectroscopic system that is developed for quantification and diagnosis of mucosa abnormalities. The system stability and reliability are firstly assessed by phantom experiments, showing a measurement variation lower than 1% within 20 min. In vitro and in vivo validations are then conducted for tissue identification and lesion differentiation. For enhanced decision, a data‐driven diagnosis algorithm is explored in pilot under different experimental configurations. The results conclude a promising accuracy of >96% for the in vivo classification as well as an excellent sensitivity of >88% for the in vitro mucosa lesions detection, and demonstrate sound potential of the system in early detection of mucosa lesions. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
1864063X
Volume :
16
Issue :
10
Database :
Complementary Index
Journal :
Journal of Biophotonics
Publication Type :
Academic Journal
Accession number :
172855191
Full Text :
https://doi.org/10.1002/jbio.202300086