X-ray System of the Proton Therapy Complex "Prometheus".

Authors :
Belikhin, M. A.
Kutlubulatov, B. A.
Zhogolev, P. B.
Zavestovskaya, I. N.
Chashurin, V. I.
Chernyaev, A. P.
Shemyakov, A. E.
Source :
Bulletin of the Lebedev Physics Institute; Aug2023, Vol. 50 Issue 8, p360-364, 5p
Publication Year :
2023

Abstract

Main elements of the X-ray system of the Prometheus proton therapy complex are described. The system is integrated into the proton synchrotron outlet socket (exit cone), and allows X-ray micrographs and cone beam tomography of the patient body area of interest. The system is successfully used in clinical practice to verify the patient position immediately before irradiation. [ABSTRACT FROM AUTHOR]