Back to Search Start Over

Flow-Based CL-SMIA for the Quantification of Protein Biomarkers from Nasal Secretions in Comparison with Sandwich ELISA.

Authors :
Neumair, Julia
Kröger, Marie
Stütz, Evamaria
Jerin, Claudia
Chaker, Adam M.
Schmidt-Weber, Carsten B.
Seidel, Michael
Source :
Biosensors (2079-6374); Jul2023, Vol. 13 Issue 7, p670, 14p
Publication Year :
2023

Abstract

Protein biomarkers in nasal secretions can be used as a measure to differentiate between allergies, airway diseases and infections for non-invasive diagnostics. The point-of-care quantification of biomarker levels using flow-based microarray facilitates precise and rapid diagnosis and displays the potential for targeted and effective treatment. For the first time, we developed a flow-based chemiluminescence sandwich microarray immunoassay (CL-SMIA) for the quantification of nasal interferon-beta (IFN-β) on the Microarray Chip Reader-Research (MCR-R). Polycarbonate foils are used as a cost-effective surface for immobilizing capture antibodies. By using a commercially available set of anti-human IFN-β antibodies, the CL-SMIA can be compared directly to an enzyme-linked immunosorbent assay (ELISA) performed in microtiter plates concerning the bioanalytical performance and economic issues. Pre-incubation of the sample with detection antibodies facilitates the lower consumption of detection antibodies, as this allows for a longer interaction time between the antibody and the biomarker. The direct injection of pre-incubated samples into the microarray chips eliminates the adsorption of proteins in the tubing as well as the contamination of the tubing and valves of the MCR-R with clinical samples. The small flow cell allows for a low sample volume of 50 μL. The limit of detection of 4.53 pg mL<superscript>−1</superscript> was slightly increased compared to a sandwich ELISA performed on microtiter plates which were 1.60 pg mL<superscript>−1</superscript>. The possibility to perform the CL-SMIA in a multiplexed mode makes it a promising assay for the rapid and cost-effective non-invasive detection of biomarkers in nasal secretions. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
20796374
Volume :
13
Issue :
7
Database :
Complementary Index
Journal :
Biosensors (2079-6374)
Publication Type :
Academic Journal
Accession number :
168598454
Full Text :
https://doi.org/10.3390/bios13070670