Class 5 integrating indicators for steam sterilization processes.
- Source :
- Healthcare Purchasing News; Apr2005, Vol. 29 Issue 4, p46-40, 3p, 2 Color Photographs, 2 Charts, 1 Graph
- Publication Year :
- 2005
-
Abstract
- A chemical integrator labeled as a Class 5 integrating indicator can be used inside each package as an internal chemical indicator to determine that the sterilant has penetrated the package. In addition, a Class 5 integrating indicator inside a PCD (process challenge pack or test pack) can be used for load control to release loads that do not contain an implantable medical device. Class 5 integrating indicators should not replace the use of biological indicators for releasing loads containing implacable, or for routine or sterilizer qualification testing, product testing or for a recall because the response of Class 3 integrating indicators does not correlate to the response of biological indicators under all sterilization process failure conditions. The selection, use and interpretation of Class 5 integrating indicator results will improve the sterilization monitoring process if you use them according to the AAMI ST46 recommended practices and understand their limitations. [ABSTRACT FROM AUTHOR]