Back to Search Start Over

Risk of magnetic resonance imaging-induced magnet dislocation for different types of cochlear implants: a single-center retrospective study.

Authors :
Rupp, Robin
Balk, Matthias
Sievert, Matti
Leibl, Victoria
Schleder, Stephan
Allner, Moritz
Iro, Heinrich
Hoppe, Ulrich
Hornung, Joachim
Gostian, Antoniu-Oreste
Source :
Journal of Otolaryngology -- Head & Neck Surgery; 4/21/2023, Vol. 52 Issue 1, p1-9, 9p
Publication Year :
2023

Abstract

Background: When performing magnetic resonance imaging (MRI) in patients with a cochlear implant (CI), complication rates vary widely in the literature. The primary objective of this retrospective study was to determine the prevalence of complications, in particular magnet dislocation, in patients with CI undergoing 1.5 Tesla (T) MRI. As a secondary objective, the prevalence of magnet dislocation for specific cochlear implant device types was elaborated. Methods: In a single-center retrospective study, all patients with a cochlear implant presenting for an MRI examination at 1.5 T at our institution between January 1st, 2010 and December 31st, 2020 were included. Implants with axial and diametrical magnets were included in the study. MRI safety measures were applied before imaging. The prevalence of complications was evaluated. Magnet dislocation rates were calculated for device types with at least 20 MRI exposures. Results: During the study period, 196 MRI examinations were performed in a total of 128 patients, accounting for 149 different implants (21 implanted bilaterally) with a total of 231 implant exposures to MRI (average 1.69 ± 1.57; min. 1, max. 12). Complications were reported in 50 out of 231 cochlear implant exposures (21.6%). Magnet dislocation occurred in a total of 27 cases (11.7%). Dislocation rates were 29.6% for the Cochlear® CI500 series (24 dislocations from 81 exposures), 1.1% for the Cochlear® CI24RE series (1 from 87) and 0% for the MED-EL® Synchrony (0 from 36). The dislocation rate for the CI500 was significantly higher than for the CI24RE (χ<superscript>2</superscript><subscript>(1)</subscript> = 26.86; p < 0.001; ϕ = 0.40) or the Synchrony (χ<superscript>2</superscript><subscript>(1)</subscript> = 13.42; p < 0.001; ϕ = 0.34). Conclusions: For 1.5 T MRI, the risk of magnet dislocation ranges from 0 to 29.6% and depends on the CI device type. Implants with a diametrical magnet can be considered potentially MRI-safe, whereas in CIs with axial magnets, the CI500 is at high risk of magnet dislocation. Therefore, apart from a strict indication for an MRI and adherence to safety protocols, post-MRI follow-up examination to rule out magnet dislocation is recommended. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
07077270
Volume :
52
Issue :
1
Database :
Complementary Index
Journal :
Journal of Otolaryngology -- Head & Neck Surgery
Publication Type :
Academic Journal
Accession number :
163252607
Full Text :