Back to Search Start Over

The Apple Watch spO2 sensor and outliers in healthy users.

Authors :
Schröder, Christina
Förster, Robert
Zwahlen, Daniel R.
Windisch, Paul
Source :
NPJ Digital Medicine; 4/8/2023, Vol. 6 Issue 1, p1-3, 3p
Publication Year :
2023

Abstract

It is unclear how frequently the Apple Watch produces spO<subscript>2</subscript> measurements outside of the normal range in healthy individuals at rest. We conducted a head-to-head comparison in 38 healthy individuals between two watches and two medical-grade pulse oximeters. Fourteen percent of watch measurements yielded spO<subscript>2</subscript> values below 95%, with no values below 92%. Results suggest that outliers measured by the watch should not be a cause for concern in otherwise healthy individuals. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
23986352
Volume :
6
Issue :
1
Database :
Complementary Index
Journal :
NPJ Digital Medicine
Publication Type :
Academic Journal
Accession number :
162970769
Full Text :
https://doi.org/10.1038/s41746-023-00814-x