Back to Search Start Over

INCREASING THE EFFICIENCY OF THE SURFACE-MOUNTED ULTRASONIC ELECTROMAGNETIC-ACOUSTIC TRANSDUCER ON ACCOUNT OF THE MAGNETIC FIELD SOURCE.

Authors :
Suchkov, G. M.
Bolyukh, V. F.
Kocherga, A. I.
Mygushchenko, R. P.
Kropachek, O. Yu.
Source :
Technical Electrodynamics / Tekhnichna Elektrodynamika; Apr2023, Issue 2, p3-8, 6p
Publication Year :
2023

Abstract

Model studies were carried out using the COMSOL Multiphysics package, aimed at ensuring the forming of a permanent magnet magnetic field at a considerable distance to a ferromagnetic product from its pole, which is necessary to create efficient portable ultrasonic electromagnetic-acoustic transducers of thickness gauges and testing and diagnostic devices. It is theoretically shown and experimentally confirmed that for portable measuring ultrasonic devices it is expedient to set the height of the permanent magnet at about 60 mm and the cross section of the magnet pole 50x50 mm². At the same time, with a gap between the magnet pole and the product of about 30 mm, the value of the normal component of the magnetic field induction near the surface of the object is about 0.3...0.4 T, which is sufficient for thickness gauging or diagnostics of ferromagnetic products using the ultrasonic pitch-and-catch method. References 19, figures 8. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
16077970
Issue :
2
Database :
Complementary Index
Journal :
Technical Electrodynamics / Tekhnichna Elektrodynamika
Publication Type :
Academic Journal
Accession number :
162247448
Full Text :
https://doi.org/10.15407/techned2023.02.003