Fast Sub-Hz potentiostatic/galvanostatic bio-impedance measurements using chaotic oscillators.
- Source :
- Biomedical Microdevices; Dec2022, Vol. 24 Issue 4, p1-11, 11p
- Publication Year :
- 2022
-
Abstract
- The measurement of bio-impedance spectra at ultra low frequencies (sub-Hz) is known to require a considerably long time with the classical frequency-sweep method or other narrow-band periodic excitation signals. In this work, an impedance measurement technique based on using wide-band chaotic signals is proposed and experimentally validated over the frequency range 10 m H z - 1 H z . The technique was tested in both potentiostatic and galvanostatic modes, first using commercial components and then using an enhanced Howland current pump designed and fabricated in a 65nm CMOS technology. The accuracy of the proposed technique was assessed on fruit samples compared to measurements conducted using a research-grade Biologic VSP-300 electro-chemical station. [ABSTRACT FROM AUTHOR]
Details
- Language :
- English
- ISSN :
- 13872176
- Volume :
- 24
- Issue :
- 4
- Database :
- Complementary Index
- Journal :
- Biomedical Microdevices
- Publication Type :
- Academic Journal
- Accession number :
- 160123570
- Full Text :
- https://doi.org/10.1007/s10544-022-00641-z