Back to Search Start Over

Micro-Volume Blood Separation Membrane for In-Situ Biosensing.

Authors :
Zhu, Qin
Wu, Huimin
Ma, Zhen
Liu, Yuqiao
Li, Junmin
Zhu, Ling
Zhang, Xinran
Wang, Chengcheng
Chen, Dajing
Zhu, Danhua
Source :
Biosensors (2079-6374); Sep2022, Vol. 12 Issue 9, p712-712, 10p
Publication Year :
2022

Abstract

In this paper, we report a point-of-care (POCT) testing strip based on a porous membrane structure for whole blood separation and colorimetric analysis without external supporting equipment. Conventional blood tests rely on large instruments for blood pretreatment and separation to improve measurement accuracy. Cellulose acetate (CA) membranes with different pore diameters and structures were prepared via a non-solvent method for the separation of whole blood. Among them, CA@PEG-2000 membranes with nano-pores on the surface and micro-pores in the interior facilitated the capture of blood cells on the surface, as well as the free diffusion of plasma through the porous interior structure. The fluid flow of blood in the asymmetric porous structure can be theoretically estimated using the Lucas-Washburn equation. Compared with the conventional paper strips and other porous membranes, the CA@PEG-2000 membrane with an immobilized sensing layer exhibited efficient blood separation, a short response time (less than 2 min), an ultralow dosage volume (5 μL), and high sensitivity. The fabricated blood separation membranes can be further used for the detection of various biomarkers in whole blood, providing additional options for rapid quantitative POCT tests. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
20796374
Volume :
12
Issue :
9
Database :
Complementary Index
Journal :
Biosensors (2079-6374)
Publication Type :
Academic Journal
Accession number :
159273121
Full Text :
https://doi.org/10.3390/bios12090712