Back to Search Start Over

Non‐destructive analysis of hollow‐shaped single fibers using X‐ray computed tomography.

Authors :
Tadano, Wataru
Takeuchi, Masaru
Tanabe, Eishi
Stellhorn, Jens R.
Komaguchi, Kenji
Nakamoto, Akihiro
Honda, Sadao
Hayakawa, Shinjiro
Source :
Journal of Forensic Sciences; Jul2022, Vol. 67 Issue 4, p1461-1467, 7p
Publication Year :
2022

Abstract

A commercial high‐resolution X‐ray computed tomography (CT) was utilized for non‐destructive analysis of single fibers. The micro‐CT apparatus was employed because it is applicable to both colored and colorless fibers. A sample preparation using adhesive sheets was demonstrated, and the method is similar to typical tape‐lift sample collection method in crime cases. Different cross‐sectional shapes of nylon and polyester single fibers were non‐destructively distinguished, and the method is applicable to all types of fibers. Cross‐sectional areas, aperture ratios, and volumes of individual fibers were directly and automatically measured using the open‐source software. The observed parameters were within a coefficient of variation of 3%. In addition, a mass of a single fragment of a fiber can be estimated when the local density is given. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
00221198
Volume :
67
Issue :
4
Database :
Complementary Index
Journal :
Journal of Forensic Sciences
Publication Type :
Academic Journal
Accession number :
157777134
Full Text :
https://doi.org/10.1111/1556-4029.15055