Fabry-Perot Interferometer Based on a Fiber-Tip Fixed-Supported Bridge for Fast Glucose Concentration Measurement.

Authors :
Tang, Shuo
Zou, Mengqiang
Zhao, Cong
Jiang, Yihang
Chen, Ribao
Xu, Zhourui
Yang, Chengbin
Wang, Xiaomei
Dong, Biqin
Wang, Yiping
Liao, Changrui
Xu, Gaixia
Source :
Biosensors (2079-6374); Jun2022, Vol. 12 Issue 6, p391-391, 11p
Publication Year :
2022

Abstract

Blood glucose concentration is important for metabolic homeostasis in humans and animals. Many diabetic patients need to detect blood glucose daily which burdens community hospitals and family healthcare. Optical fiber sensors are widely used in biomedical detection because of their compact structure, fast response, high sensitivity, low cost, and ease of operation. In this work, we constructed a Fabry-Perot (FP) cavity biosensor for the fast detection of glucose concentration in serum. The femtosecond laser micromachining was applied to fabricate the FP cavity by printing the fiber-tip fixed-supported bridge at the end face of the optical fiber. An additional hemisphere was printed at the center of the outer surface of the bridge to avoid multi-beam interference. The results demonstrated that the proposed biosensor had high refractive index (RI) detection sensitivity, roughly 1039 nm/RIU at a wavelength of 1590 nm, and the detection sensitivity for glucose was around 0.185 nm/ (mg/mL) at a wavelength of 1590 nm. Due to its high sensitivity, compact structure, and fast response, the FP cavity biosensor has great potential to be applied in family healthcare for glucose concentration detection of diabetic patients. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
20796374
Volume :
12
Issue :
6
Database :
Complementary Index
Journal :
Biosensors (2079-6374)
Publication Type :
Academic Journal
Accession number :
157661299
Full Text :
https://doi.org/10.3390/bios12060391