Fiducial marker position affects target volume in stereotactic lung irradiation.

Authors :
Akasaka, Hiroaki
Mizonobe, Kazufusa
Oki, Yuya
Uehara, Kazuyuki
Nakayama, Masao
Tamura, Shuhei
Munetomo, Yoshiki
Kawaguchi, Haruna
Ishida, Jun
Harada, Aya
Ishihara, Takeaki
Kubota, Hikaru
Kawaguchi, Hiroki
Sasaki, Ryohei
Mayahara, Hiroshi
Source :
Journal of Applied Clinical Medical Physics; Jun2022, Vol. 23 Issue 6, p1-10, 10p
Publication Year :
2022

Abstract

Purpose: Real‐time tracking systems of moving respiratory targets such as CyberKnife, Radixact, or Vero4DRT are an advanced robotic radiotherapy device used to deliver stereotactic body radiotherapy (SBRT). The internal target volume (ITV) of lung tumors is assessed through a fiducial marker fusion using four‐dimensional computed tomography (CT). It is important to minimize the ITV to protect normal lung tissue from exposure to radiation and the associated side effects post SBRT. However, the ITV may alter if there is a change in the position of the fiducial marker with respect to the tumor. This study investigated the relationship between fiducial marker position and the ITV in order to prevent radiation exposure of normal lung tissue, and correct target coverage. Materials and methods: This study retrospectively reviewed 230 lung cancer patients who received a fiducial marker for SBRT between April 2015 and September 2021. The distance of the fiducial marker to the gross tumor volume (GTV) in the expiratory (dex) and inspiratory (din) CT, and the ratio of the ITV/V(GTVex), were investigated. Results: Upon comparing each lobe, although there was no significant difference in the ddiff and the ITV/V(GTVex) between all lobes for dex < 10 mm, there was significant difference in the ddiff and the ITV/V(GTVex) between the lower and upper lobes for dex ≥ 10 mm (p < 0.05). Moreover, there was significant difference in the ddiff and the ITV/V(GTVex) between dex ≥10 mm and dex < 10 mm in all lung regions (p < 0.05). Conclusion: The ITV that had no margin from GTVs increased when dex was ≥10 mm for all lung regions (p < 0.05). Furthermore, the increase in ITV tended to be greater in the lower lung lobe. These findings can help decrease the possibility of adverse events post SBRT, and correct target coverage. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
15269914
Volume :
23
Issue :
6
Database :
Complementary Index
Journal :
Journal of Applied Clinical Medical Physics
Publication Type :
Academic Journal
Accession number :
157443335
Full Text :
https://doi.org/10.1002/acm2.13596