Back to Search
Start Over
Agreement of Tear Break-Up Time and Meniscus Height between Medmont E300 and Visionix VX120+.
- Source :
- Applied Sciences (2076-3417); May2022, Vol. 12 Issue 9, pN.PAG-N.PAG, 8p
- Publication Year :
- 2022
-
Abstract
- The goal of this study was to analyze the agreement between the Medmont E300 and the Visionix VX120+ systems in terms of non-invasive tear break-up time (NIBUT) and tear meniscus height (TMH) measurements. A total of 60 eyes (30 healthy subjects) were enrolled. NIBUT and TMH were evaluated with Medmont E300; first NIBUT, NIBUT50%, and TMH were evaluated with Visionix VX120+. Both evaluations were performed in a random order by the same clinician for right, left, and both eyes. The Medmont E300 provided significantly higher NIBUT than Visionix VX120+ for first NIBUT in right, left, and both eyes (p ≤ 0.003) and NIBUT50% in left and both eyes (p ≤ 0.042). The TMH measured with VX120+ was significantly higher than with Medmont E300 considering both eyes (p = 0.037). No significant correlations were found between both devices for either NIBUT (p ≥ 0.11) or TMH (p ≥ 0.09). Passing–Bablok regression analyses revealed poor agreement between devices for NIBUT and TMH outcomes. VX120+ is expected to provide substantial lower first NIBUT values than the NIBUT measured by Medmont E300. Clinicians should consider not using both instruments as interchangeable for dry eye diagnosis. [ABSTRACT FROM AUTHOR]
- Subjects :
- DRY eye syndromes
REGRESSION analysis
MEDICAL personnel
Subjects
Details
- Language :
- English
- ISSN :
- 20763417
- Volume :
- 12
- Issue :
- 9
- Database :
- Complementary Index
- Journal :
- Applied Sciences (2076-3417)
- Publication Type :
- Academic Journal
- Accession number :
- 156850348
- Full Text :
- https://doi.org/10.3390/app12094589