Design and Evaluation of MeVisLab Networks for Co-Registration and Cropping of Positron Emission Tomography/Computed Tomography Scans.

Authors :
Pandey, Anil Kumar
Sharma, Akshima
Kumar, Sunil
Negi, Ashish
Singh, Satnam
Kumar, Sanjay
Kumar, Rakesh
Source :
Indian Journal of Nuclear Medicine; Jan-Mar2022, Vol. 37 Issue 1, p37-42, 6p
Publication Year :
2022

Abstract

Objective: The aim of the present study was to design and evaluate two MeVisLab networks, one for co-registration of positron emission tomography/computed tomography (PET/CT) images and second for cropping the co-registered PET/CT images. Materials and Methods: Two MeVisLab networks, one to co-register and export PET/CT DICOM images and second for cropping the co-registered PET/CT images were designed using different modules of registration toolkit MERIT. One hundred and twenty-five PET/CT studies were exported from Siemens and GE scanners in DICOM format. These images were co-registered and cropped with our designed networks. The images co-registered with our network were compared visually with the co-registered images of same PET/CT studies on vendor provided workstations by an experienced nuclear medicine physician (NMP). The perfection of the cropping of co-registered images was also assessed visually. Results: Visually, NMP found all 125 images co-registered using the network designed in our study similar to the co-registered images of vendor provided workstations. Furthermore, the cropping of all co-registered images was perfectly done by our network. Conclusion: Two MeVisLab networks designed and evaluated in the present study can be used for co-registration of PET/CT DICOM images and cropping the co-registered PET/CT images. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
09723919
Volume :
37
Issue :
1
Database :
Complementary Index
Journal :
Indian Journal of Nuclear Medicine
Publication Type :
Academic Journal
Accession number :
156044623
Full Text :
https://doi.org/10.4103/ijnm.ijnm_107_21