Back to Search
Start Over
Magnetic, Mechanical and Thermal Modeling of Superconducting, Whole-Body, Actively Shielded, 3 T MRI Magnets Wound Using MgB 2 Strands for Liquid Cryogen Free Operation.
- Source :
- IEEE Transactions on Applied Superconductivity; Sep2022, Vol. 32 Issue 6, p1-4, 4p
- Publication Year :
- 2022
-
Abstract
- We present magnetic, mechanical and thermal modeling results for a 3 Tesla actively shielded whole body MRI (Magnetic Resonance Imaging) magnet consisting of coils with a square cross section of their windings. The magnet design was a segmented coil type optimized to minimize conductor length while hitting the standard field quality and DSV (Diameter of Spherical Volume) specifications as well as a standard, compact size 3 T system. It had an overall magnet length and conductor length which can lead to conduction cooled designs comparable to NbTi helium bath cooled 3 T MRI magnets. The design had a magnetic field homogeneity better than 10 ppm (part-per-million) within a DSV (Diameter of Spherical Volume) of 48 cm and the total magnet winding length of 1.37 m. A new class of MgB2 strand especially designed for MRI applications was considered as a possible candidate for winding such magnets. This work represents the first magnetic, mechanical and thermal design for a whole-body 3 T MgB2 short (1.37 m length) MRI magnet based on the performance parameters of existing MgB2 wire. 3 Tesla MRI magnet can operate at 20 K at 67% of its critical current. [ABSTRACT FROM AUTHOR]
Details
- Language :
- English
- ISSN :
- 10518223
- Volume :
- 32
- Issue :
- 6
- Database :
- Complementary Index
- Journal :
- IEEE Transactions on Applied Superconductivity
- Publication Type :
- Academic Journal
- Accession number :
- 155601957
- Full Text :
- https://doi.org/10.1109/TASC.2022.3147137