Lead extractions: dissecting adhesions up to the lead‐tip of the right ventricle: safety and success‐rates.

Authors :
Witte, Olivier A.
Delnoy, Peter Paul HM
Ghani, Abdul
Smit, Jaap Jan J.
Ramdat Misier, Anand R.
Elvan, Arif
Adiyaman, Ahmet
Source :
Pacing & Clinical Electrophysiology; Jan2022, Vol. 45 Issue 1, p132-140, 9p
Publication Year :
2022

Abstract

Aims: Goal of Transvenous Lead Extraction (TLE) is complete removal of all targeted leads, without complications. Despite counter traction manoeuvres, efficacy rates are often hampered by broken right ventricle lead (RV‐lead) tips. Mechanically powered lead extraction (Evolution sheath) is effective, however safety of dissection up to the lead tip is unclear. Therefore, we examined the feasibility and safety of RV‐lead extraction requiring dissection up to the myocardium. Methods and results: From 2009 to 2018, all TLE in the Isala Heart Centre (Zwolle, The Netherlands) requiring the hand‐powered mechanical Evolution system to extract RV‐leads (n = 185) were examined from a prospective registry. We assessed 4 groups: TLE with the first generation Evolution (n = 43) with (A1,n = 18) and without (A2,n = 25) adhesions up to the myocardium and TLE with the Novel R/L type (n = 142) of sheath with (B1, n = 59) and without (B2, n = 83) adhesions up to the myocardium. Complete success rate in Group B was significantly higher than group A (96.5 vs 76.7%, p = 0.0354). When comparing the patients with adhesions up to the myocardium, total complete success is higher in the R/L group (61.1% vs 90.5%, p = 0.0067). There were no deaths. Overall major complication rates were low (2/185; 1.1%) and there was no statistically significant difference in major and minor complications between the two groups. Conclusion: Extraction strategy with the bidirectional Evolution R/L sheath for right ventricular leads with adhesions up to the myocardium is safe and feasible. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
01478389
Volume :
45
Issue :
1
Database :
Complementary Index
Journal :
Pacing & Clinical Electrophysiology
Publication Type :
Academic Journal
Accession number :
154741116
Full Text :
https://doi.org/10.1111/pace.14416