Back to Search Start Over

Renal hemofiltration prevents metabolic acidosis and reduces inflammation during normothermic machine perfusion of the vascularized composite allograft: A preclinical study.

Authors :
Stone, John P.
Amin, Kavit R.
Geraghty, Abbey
Kerr, Jak
Shaw, Matthew
Dabare, Dilan
Wong, Jason K.
Brough, David
Entwistle, Timothy R.
Montero‐Fernandez, Angeles
Fildes, James E.
Source :
Artificial Organs; Feb2022, Vol. 46 Issue 2, p259-272, 14p
Publication Year :
2022

Abstract

Introduction: Recent experimental evidence suggests normothermic machine perfusion of the vascularized composite allograft results in improved preservation compared to static cold storage, with less reperfusion injury in the immediate post‐operative period. However, metabolic acidosis is a common feature of vascularized composite allograft perfusion, primarily due to the inability to process metabolic by‐products. We evaluated the impact of combined limb‐kidney perfusion on markers of metabolic acidosis and inflammation in a porcine model. Methods: Ten paired pig forelimbs were used for this study, grouped as either limb‐only (LO, n = 5) perfusion, or limb‐kidney (LK, n = 5) perfusion. Infrared thermal imaging was used to determine homogeneity of perfusion. Lactate, bicarbonate, base, pH, and electrolytes, along with an inflammatory profile generated via the quantification of cytokines and cell‐free DNA in the perfusate were recorded. Results: The addition of a kidney to a limb perfusion circuit resulted in the rapid stabilization of lactate, bicarbonate, base, and pH. Conversely, the LO circuit became progressively acidotic, correlating in a significant increase in pro‐inflammatory cytokines. Global perfusion across the limb was more homogenous with LK compared to LO. Conclusion: The addition of a kidney during limb perfusion results in significant improvements in perfusate biochemistry, with no evidence of metabolic acidosis. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
0160564X
Volume :
46
Issue :
2
Database :
Complementary Index
Journal :
Artificial Organs
Publication Type :
Academic Journal
Accession number :
154612334
Full Text :
https://doi.org/10.1111/aor.14089