Back to Search Start Over

Comparison of electrocardiograms (ECG) waveforms and centralized ECG measurements between a simple 6-lead mobile ECG device and a standard 12-lead ECG.

Authors :
Kleiman, Robert
Darpo, Borje
Brown, Randy
Rudo, Todd
Chamoun, Svetlana
Albert, David E.
Bos, Johan Martijn
Ackerman, Michael J.
Source :
Annals of Noninvasive Electrocardiology; Nov2021, Vol. 26 Issue 6, p1-10, 10p
Publication Year :
2021

Abstract

Background: Interval duration measurements (IDMs) were compared between standard 12-lead electrocardiograms (ECGs) and 6-lead ECGs recorded with AliveCor's KardiaMobile 6L, a hand-held mobile device designed for use by patients at home. Methods: Electrocardiograms were recorded within, on average, 15 min from 705 patients in Mayo Clinic's Windland Smith Rice Genetic Heart Rhythm Clinic. Interpretable 12-lead and 6-lead recordings were available for 685 out of 705 (97%) eligible patients. The most common diagnosis was congenital long QT syndrome (LQTS, 343/685 [50%]), followed by unaffected relatives and patients (146/685 [21%]), and patients with other genetic heart diseases, including hypertrophic cardiomyopathy (36 [5.2%]), arrhythmogenic cardiomyopathy (23 [3.4%]), and idiopathic ventricular fibrillation (14 [2.0%]). IDMs were performed by a central ECG laboratory using lead II with a semi-automated technique. Results: Despite differences in patient position (supine for 12-lead ECGs and sitting for 6-lead ECGs), mean IDMs were comparable, with mean values for the 12-lead and 6-lead ECGs for QTcF, heart rate, PR, and QRS differing by 2.6 ms, -5.5 beats per minute, 1.0 and 1.2 ms, respectively. Despite a modest difference in heart rate, intervals were close enough to allow a detection of clinically meaningful abnormalities. Conclusions: The 6-lead hand-held device is potentially useful for a clinical follow-up of remote patients, and for a safety follow-up of patients participating in clinical trials who cannot visit the investigational site. This technology may extend the use of 12-lead ECG recordings during the current COVID-19 pandemic as remote patient monitoring becomes more common in virtual or hybrid-design clinical studies. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
1082720X
Volume :
26
Issue :
6
Database :
Complementary Index
Journal :
Annals of Noninvasive Electrocardiology
Publication Type :
Academic Journal
Accession number :
154231172
Full Text :
https://doi.org/10.1111/anec.12872