Feasibility of a Drug-Releasing Radiofrequency Ablation System in a Porcine Liver Model.

Authors :
Cho, Young Chul
Lee, Ki Baek
Ham, Su Jung
Jung, Jin Hwa
Park, Yubeen
Won, Dong-Sung
Kim, Kyung Won
Park, Jung-Hoon
Source :
Applied Sciences (2076-3417); Sep2021, Vol. 11 Issue 18, p8301, 11p
Publication Year :
2021

Abstract

The objective of this study was to investigate the feasibility of a newly developed anticancer drug-releasing radiofrequency ablation (RFA) system in a porcine liver model. A 15-gauge drug-releasing cooled wet electrode (DRCWE) was newly developed to improve the RFA efficacy for creating a large ablation as well as for simultaneously delivering an anticancer drug to the tumor margin. Nine ablations in three pigs were performed by the DRCWE. The sectioned liver specimens were evaluated by measuring the ablation zone by a positron emission tomography/magnetic resonance imaging examination to investigate whether <superscript>18</superscript>F-fluorodeoxyglucose was exactly diffused. Volumes of the ablation zones released drug injection volumes, circularity, retention rate defined as the ratio between an estimated and injection dose, and the standard uptake value were assessed. The drug-releasing RFA was technically successful without procedural-related complications. During the procedure, the color changes of the ablated zones of the liver were observed in all specimens. The mean drug injection volume was higher than the ablated volumes (17.21 ± 2.85 vs. 15.22 ± 2.30 cm<superscript>3</superscript>) and the circularity was 0.72 ± 0.08. Moreover, the retention rate was 72.89% ± 4.22% and the mean standard uptake value was 0.44 ± 0.05. The drug-releasing RFA system was feasible not only for local ablation but also for the delivery of anticancer drugs. The results of this study indicate that this novel strategy of localized RFA with a drug delivery system could be a promising option for the prevention of local recurrence rates. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
20763417
Volume :
11
Issue :
18
Database :
Complementary Index
Journal :
Applied Sciences (2076-3417)
Publication Type :
Academic Journal
Accession number :
152657399
Full Text :
https://doi.org/10.3390/app11188301