Back to Search
Start Over
Characterizing x-ray transmission through filters used in high energy density physics diagnostics.
- Source :
- Review of Scientific Instruments; 2021, Vol. 92 Issue 6, p1-5, 5p
- Publication Year :
- 2021
-
Abstract
- We report on the design and implementation of a new system used to characterize the energy-dependent x-ray transmission curve, Θ(E), through filters used in high-energy density physics diagnostics. Using an Amptek X-123-CdTe x-ray spectrometer together with a partially depleted silicon surface barrier detector, both the energy spectrum and total emission of an x-ray source have been accurately measured. By coupling these detectors with a custom PROTO-XRD x-ray source with interchangeable cathodes, accurate characterizations of Θ(E) for filters of varying materials and thicknesses have been obtained. The validity of the technique has been confirmed by accurately reproducing areal densities for high-purity filters with known x-ray transmission properties. In this paper, the experimental setup is described and the results of absorption calibrations performed on a variety of different filters are presented. [ABSTRACT FROM AUTHOR]
Details
- Language :
- English
- ISSN :
- 00346748
- Volume :
- 92
- Issue :
- 6
- Database :
- Complementary Index
- Journal :
- Review of Scientific Instruments
- Publication Type :
- Academic Journal
- Accession number :
- 151193506
- Full Text :
- https://doi.org/10.1063/5.0043770