Micro-CT Analysis of Microgap at a Novel Two-Piece Dental Implant Comprising a Replaceable Sleeve In Vitro.

Authors :
Kapishnikov, Sergey
Gadyukov, Adir
Chaushu, Gavriel
Chaushu, Liat
Source :
International Journal of Oral & Maxillofacial Implants; May/Jun2021, Vol. 36 Issue 3, p451-459, 9p, 5 Color Photographs, 3 Charts, 2 Graphs
Publication Year :
2021

Abstract

Microcomputed tomography (micro-CT) is a relatively new modality to investigate mechanical deformations. The purpose of this study was to assess the microgap at the implant-sleeve connection of a new two-piece dental implant with a replaceable sleeve. Materials and Methods: Implants were assembled with 25-degree angulated abutments. Micro-CT was used to assess implant-sleeve connection gaps under the following mechanical conditions: (1) unloading; (2) compressive 10,000 cyclic loading with 400 N; (3) static compressive load of 200 N or 400 N for 24 hours. Results: The mean gap in the unloaded sample was 2.9 ± 0.9 μm. The mean gap difference after cyclic compressive load was 0.3 ± 0.15 μm, demonstrating a negligible effect for the cyclic loading. Under static compressive load, there was no increase in microgap size at 200 N. At 400 N, a significant (P < .05) increase was noted. While the mean values increased by 1.9 μm, the most pronounced significant increase in mean microgap was noted in the direction of force application (5.1 ± 2.14 μm), while a significant decrease in mean microgap (1.2 ± 1.47 μm) was noted on the opposite side. Conclusion: The mechanical behavior of the implant-sleeve connection under static and dynamic loads was found to be within the previously reported range of implant dentistry. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
08822786
Volume :
36
Issue :
3
Database :
Complementary Index
Journal :
International Journal of Oral & Maxillofacial Implants
Publication Type :
Academic Journal
Accession number :
150914715
Full Text :
https://doi.org/10.11607/jomi.8563