Application of lung substitute material as ripple filter for multi-ion therapy with helium-, carbon-, oxygen-, and neon-ion beams.

Authors :
Inaniwa, Taku
Abe, Yasushi
Suzuki, Masao
Lee, Sung Hyun
Mizushima, Kota
Nakaji, Taku
Sakata, Dousatsu
Sato, Shinji
Iwata, Yoshiyuki
Kanematsu, Nobuyuki
Shirai, Toshiyuki
Source :
Physics in Medicine & Biology; 3/7/2021, Vol. 66 Issue 5, p1-10, 10p
Publication Year :
2021

Abstract

A development project for hypo-fractionated multi-ion therapy has been initiated at the National Institute of Radiological Sciences in Japan. In the treatment, helium, carbon, oxygen, and neon ions will be used as primary beams with pencil beam scanning. A ripple filter (RiFi), consisting of a thin plastic or aluminum plate with a fine periodic ridge and groove structure, has been used to broaden the Bragg peak of heavy-ion beams in the beam direction. To sufficiently broaden the Bragg peak of helium-, carbon-, oxygen-, and neon-ion beams with suppressed lateral scattering and surface dose inhomogeneity, in this study, we tested a plate made of a lung substitute material, Gammex LN300, as the RiFi. The planar integrated dose distribution of a 183.5 MeV u<superscript>−1</superscript> neon-ion beam was measured behind a 3 cm thick LN300 plate in water. The Bragg peak of the pristine beam was broadened following the normal distribution with the standard deviation σ value of 1.29 mm, while the range of the beam was reduced by 8.8 mm by the plate. To verify the LN300 performance as the RiFi in multi-ion therapy, we measured the pencil beam data of helium-, carbon-, oxygen- and neon-ion beams penetrating the 3 cm thick LN300 plate. The data were then modeled and used in a treatment planning system to achieve a uniform 10% survival of human undifferentiated carcinoma cells within a cuboid target by the beam for each of the different ion species. The measured survival fractions were reasonably reproduced by the planned ones for all the ion species. No surface dose inhomogeneity was observed for any ion species even when the plate was placed close to the phantom surface. The plate made of lung substitute material, Gammex LN300, is applicable as the RiFi in multi-ion therapy with helium-, carbon-, oxygen- and neon-ion beams. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
00319155
Volume :
66
Issue :
5
Database :
Complementary Index
Journal :
Physics in Medicine & Biology
Publication Type :
Academic Journal
Accession number :
149393527
Full Text :
https://doi.org/10.1088/1361-6560/abde99