Unlocking a closed system: dosimetric commissioning of a ring gantry linear accelerator in a multivendor environment.

Authors :
Saini, Amarjit
Tichacek, Chris
Johansson, William
Redler, Gage
Zhang, Geoffrey
Moros, Eduardo G.
Qayyum, Muqeem
Feygelman, Vladimir
Source :
Journal of Applied Clinical Medical Physics; Feb2021, Vol. 22 Issue 2, p21-34, 14p
Publication Year :
2021

Abstract

The Halcyon™ platform is self‐contained, combining a treatment planning (Eclipse) system TPS) with information management and radiation delivery components. The standard TPS beam model is configured and locked down by the vendor. A portal dosimetry‐based system for patient‐specific QA (PSQA) is also included. While ensuring consistency across the user base, this closed model may not be optimal for every department. We set out to commission independent TPS (RayStation 9B, RaySearch Laboratories) and PSQA (PerFraction, Sun Nuclear Corp.) systems for use with the Halcyon linac. The output factors and PDDs for very small fields (0.5 × 0.5 cm2) were collected to augment the standard Varian dataset. The MLC leaf‐end parameters were estimated based on the various static and dynamic tests with simple model fields and honed by minimizing the mean and standard deviation of dose difference between the ion chamber measurements and RayStation Monte Carlo calculations for 15 VMAT and IMRT test plans. Two chamber measurements were taken per plan, in the high (isocenter) and lower dose regions. The ratio of low to high doses ranged from 0.4 to 0.8. All percent dose differences were expressed relative to the local dose. The mean error was 0.0 ± 1.1% (TG119‐style confidence limit ± 2%). Gamma analysis with the helical diode array using the standard 3%Global/2mm criteria resulted in the average passing rate of 99.3 ± 0.5% (confidence limit 98.3%–100%). The average local dose error for all detectors across all plans was 0.2% ± 5.3%. The ion chamber results compared favorably with our recalculation with Eclipse and PerFraction, as well as with several published Eclipse reports. Dose distribution gamma analysis comparisons between RayStation and PerFraction with 2%Local/2mm criteria yielded an average passing rate of 98.5% ± 0.8% (confidence limit 96.9%–100%). It is feasible to use the Halcyon accelerator with independent planning and verification systems without sacrificing dosimetric accuracy. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
15269914
Volume :
22
Issue :
2
Database :
Complementary Index
Journal :
Journal of Applied Clinical Medical Physics
Publication Type :
Academic Journal
Accession number :
148722469
Full Text :
https://doi.org/10.1002/acm2.13116