High resolution transmission grating spectrometer for edge toroidal rotation measurements of tokamak plasmas.

Authors :
Graf, A.
May, M.
Beiersdorfer, P.
Magee, E.
Lawrence, M.
Rice, J.
Source :
Review of Scientific Instruments; Oct2004 Part I & II, Vol. 75 Issue 10, p4165-4167, 3p
Publication Year :
2004

Abstract

We present a high throughput (f/3) visible (3500–7000 Å) Doppler spectrometer for toroidal rotation velocity measurements of the Alcator C-Mod tokamak plasma. The spectrometer has a temporal response of 1 ms and a rotation velocity sensitivity of ∼10<superscript>5</superscript> cm/s. This diagnostic will have a tangential view and map out the plasma rotation at several locations along the outer half of the minor radius (r/a>0.5). The plasma rotation will be determined from the Doppler shifted wavelengths of D<subscript>α</subscript> and magnetic and electric dipole transitions of highly ionized impurities in the plasma. The fast time resolution and high spectral resolving power are possible due to a 6 in. diam circular transmission grating that is capable of λ/Δλ∼15 500 at 5769 Å in conjunction with a 50 μm slit. [ABSTRACT FROM AUTHOR]