Separatrix radius measurement of field-reversed configuration plasma in FRX-L.
- Source :
- Review of Scientific Instruments; Oct2004 Part I & II, Vol. 75 Issue 10, p4289-4292, 4p
- Publication Year :
- 2004
-
Abstract
- Magnetic pickup coils and single turn flux loops are installed on the FRX-L device. The combination of the two measurements provides the excluded flux radius that approximates the separatrix radius of the field-reversed configuration (FRC) plasma. Arrays of similar probes are used to map out local magnetic field dynamics beyond both ends of the theta-coil confinement region to help understand the effects of cusp locations on flux trapping during the FRC formation process. Details on the probe design and system calibrations are presented. The overall system calibration of excluded flux radius measurement is examined by replacing FRC plasma with a known radius aluminum conductor cylinder. [ABSTRACT FROM AUTHOR]
- Subjects :
- PHYSICS instruments
PLASMA gases
RADIAL bone
MAGNETIC fields
ALUMINUM
CALIBRATION
Subjects
Details
- Language :
- English
- ISSN :
- 00346748
- Volume :
- 75
- Issue :
- 10
- Database :
- Complementary Index
- Journal :
- Review of Scientific Instruments
- Publication Type :
- Academic Journal
- Accession number :
- 14842079
- Full Text :
- https://doi.org/10.1063/1.1789618