Back to Search Start Over

In-situ calibration technology of liquid level with precise laser ranging.

Authors :
Fan, Baixing
Zhou, Weihu
Luan, Hao
Xiang, Minzhi
Wang, Deli
Source :
Journal of Engineering; Dec2019, Vol. 2019 Issue 12, p8839-8842, 4p
Publication Year :
2019

Abstract

Here, in order to calibrate the magnetostrictive liquid-level meter in situ, a set of in-situ calibration device for magnetostrictive liquid-level meter is designed. The structure and working principle of the device are described, and the experimental test and error analysis are carried out for the parts of the in-situ calibration device. Finally, according to the modern instrument and error theory, the accuracy of the calibration device is assessed, and the field calibration experiment of magnetostrictive liquid level gauge is carried out. The theoretical analysis and field experiment results show that the calibrated device can meet the in-situ calibration requirements of the magnetostrictive liquid-level meter with a range of 0–6 m and a precision of 1 mm. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
20513305
Volume :
2019
Issue :
12
Database :
Complementary Index
Journal :
Journal of Engineering
Publication Type :
Academic Journal
Accession number :
148149510
Full Text :
https://doi.org/10.1049/joe.2018.9120