Back to Search
Start Over
Enhanced spatial resolution of Eljen-204 plastic scintillators for use in rep-rated proton diagnostics.
- Source :
- Review of Scientific Instruments; Oct2020, Vol. 91 Issue 10, p1-7, 7p
- Publication Year :
- 2020
-
Abstract
- A pixelated scintillator has been designed, fabricated, and tested using a laser-accelerated proton source for use in proton diagnostics at rep-rated laser facilities. The work presented here demonstrates the enhanced spatial resolution of thin, organic scintillators through a novel pixelation technique. Experimental measurements using laser-generated protons incident onto 130 μm-thick scintillators indicate a >20% reduction in the scintillator point spread function (PSF) for the detectors tested. The best performing pixelated detector reduced the ∼200 μm PSF of the stock material to ∼150 μm. The fabrication technique may be tailored to reduce the pixel size and achieve higher spatial resolutions. [ABSTRACT FROM AUTHOR]
- Subjects :
- PROTONS
SCINTILLATORS
DETECTORS
LASERS
PIXELS
Subjects
Details
- Language :
- English
- ISSN :
- 00346748
- Volume :
- 91
- Issue :
- 10
- Database :
- Complementary Index
- Journal :
- Review of Scientific Instruments
- Publication Type :
- Academic Journal
- Accession number :
- 146790085
- Full Text :
- https://doi.org/10.1063/5.0014949