Back to Search Start Over

Optimal echo times for multi‐gradient echo‐based B0 field‐mapping.

Authors :
Geiger, Yasmin
Tal, Assaf
Source :
NMR in Biomedicine; Jul2020, Vol. 33 Issue 7, p1-17, 17p
Publication Year :
2020

Abstract

B0 field maps are used ubiquitously in neuroimaging, in disciplines ranging from magnetic resonance spectroscopy to temperature mapping and susceptibility‐weighted imaging. Most B0 maps are acquired using standard gradient‐echo–based vendor‐provided sequences, often comprised of two echoes spaced a few milliseconds apart. Herein, we analyze the optimal spacing of echo times, defined as those maximizing precision—minimizing the standard deviation—for a fixed total acquisition time. Field estimation is carried out using a weighted least squares estimator. The standard deviation is shown to be approximately inversely proportional to the total acquisition time, suggesting a law of diminishing returns, whereby substantial gains are obtained up to a certain point, with little improvement beyond that point. Validations are provided in a phantom and a group of volunteers. Multi‐gradient echo sequences are readily available on all manufacturer platforms, making our recommendations straightforward to implement on any modern scanner. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
09523480
Volume :
33
Issue :
7
Database :
Complementary Index
Journal :
NMR in Biomedicine
Publication Type :
Academic Journal
Accession number :
143570710
Full Text :
https://doi.org/10.1002/nbm.4316