Back to Search Start Over

A robust and novel electrical impedance metric of pulmonary function in ALS patients.

Authors :
Munir, Badria
Murphy, Ethan K
Mallick, Akashleena
Gutierrez, Hilda
Zhang, Fu
Verga, Sarah
Smith, Christy
Levy, Sean
McIlduff, Courtney
Sarbesh, Pandeya
Halter, Ryan J
Rutkove, Seward B
Source :
Physiological Measurement; Apr2020, Vol. 41 Issue 4, p1-10, 10p
Publication Year :
2020

Abstract

Objective: Pulmonary function tests (PFTs) are important for assessing respiratory function in amyotrophic lateral sclerosis (ALS) patients. However, weakness of oral and glottal closure, due to concomitant bulbar dysfunction, may result in unreliable PFT values stemming from leakage of air around the breathing tube and through the glottis. In this study, we assessed whether standard thoracic electrical impedance tomography (EIT) could serve as a surrogate measure for PFTs. Approach: Thoracic EIT was performed simultaneously with standard PFTs on seven ALS patients without clinical bulbar weakness (six men and one woman, mean age of 63 years) and ten healthy volunteers (seven men and three women, mean age of 57 years). A raw impedance metric along with more standard EIT measures were computed and correlated with the normalized forced vital capacity (FVC). Additionally, test/re-test metrics and EIT images were analyzed. Main results: The impedance metric was found to be robust and sensitive to lung activity. We also identified qualitative EIT differences between healthy volunteers and ALS patients, with the ALS images showing greater heterogeneity. Significant correlations with FVC were found for both impedance and EIT metrics in ALS patients (r<superscript>2</superscript> = 0.89) and for the impedance metric only in healthy volunteers (r<superscript>2</superscript> = 0.49). Significance: This suggests that EIT, using our novel impedance metric, has the potential to serve as an alternative technology to standard PFTs for assessing pulmonary function in patients with ALS, offering new metrics of disease status for those with bulbar weakness. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
09673334
Volume :
41
Issue :
4
Database :
Complementary Index
Journal :
Physiological Measurement
Publication Type :
Academic Journal
Accession number :
143366528
Full Text :