Back to Search Start Over

High-count-rate 3He position-sensitive detector system: NEUNET-HCR.

Authors :
Ott, F.
Menelle, A.
Alba-Simionesco, C.
Sato, Setsuo
Seya, Tomohiro
Oshita, Hidetoshi
Kato, Hideto
Hikida, Noriyuki
Ishizawa, Kazuya
Yamaguchi, Akinori
Matsuura, Mitsugu
Source :
EPJ Web of Conferences; 3/11/2020, Vol. 231, p1-5, 5p
Publication Year :
2020

Abstract

Highly reliable <superscript>3</superscript>He gas detectors are most often used in neutron scattering experiments but have disadvantages such as low count rates and low position resolutions. In this study, a <superscript>3</superscript>He gas detector system called NEUNET-HCR was developed, whose count rate is approximately 10 times that of the current NEUNET system, which was developed for a high position resolution with a low count rate. NEUNET-HCR achieved a maximum count rate of 535 kcps per detector, with a loss of approximately 30%. As the NEUNET-HCR system was developed based on the NEUNET system, various improvements were required to achieve the high count rate. Further, a high count rate was achieved in the NEUNET-HCR system without significantly degrading the position resolution. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
21016275
Volume :
231
Database :
Complementary Index
Journal :
EPJ Web of Conferences
Publication Type :
Conference
Accession number :
142673058
Full Text :
https://doi.org/10.1051/epjconf/202023105004