Neutron beam performance of iBNCT as linac-based neutron source for boron neutron capture therapy.
- Source :
- EPJ Web of Conferences; 3/11/2020, Vol. 231, p1-4, 4p
- Publication Year :
- 2020
-
Abstract
- The linac-based neutron source "iBNCT" developed by the Tsukuba team has been shown to generate a large intensity of neutrons. To confirm the applicability of the device to BNCT, several characteristic measurements have been implemented. In a water phantom experiment, when the accelerator was operated with an average current of 1.4 mA, the maximum thermal neutron flux was approximately 7.8 × 10<superscript>8</superscript> (cm<superscript>-2</superscript>s<superscript>-1</superscript>). From estimation of the stability of the linac, the results demonstrate its applicability to actual treatment. [ABSTRACT FROM AUTHOR]