Back to Search Start Over

A Cryogenic Front-End Preamplifier Operating at 120 K for Bolometric Detector.

Authors :
Reza, A.
Vatsa, V.
Pose, M. S.
Mazumdar, A.
Garai, A.
Krishnamoorthy, H.
Gupta, G.
Nanal, V.
Pillay, R. G.
Shrivastava, A.
Source :
Journal of Low Temperature Physics; Apr2020, Vol. 199 Issue 1/2, p200-205, 6p
Publication Year :
2020

Abstract

In this paper, the design and test results of a front-end cryogenic amplifier for a bolometer detector are presented. The amplifier is implemented with two low-noise Si JFETs, where the first FET is configured as a source follower amplifier and the second FET provides a high dynamic resistance to keep the voltage gain very close to unity. A 3 dB bandwidth over a wide range from DC to 10 MHz is achieved in this design. The input voltage noise density is ~ 1.9 nV/√Hz at 300 K, which further reduces to 1.6 nV/√Hz at 120 K. The flicker corner frequency observed is below 50 Hz. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
00222291
Volume :
199
Issue :
1/2
Database :
Complementary Index
Journal :
Journal of Low Temperature Physics
Publication Type :
Academic Journal
Accession number :
142665196
Full Text :
https://doi.org/10.1007/s10909-019-02318-z