A Cryogenic Front-End Preamplifier Operating at 120 K for Bolometric Detector.
- Source :
- Journal of Low Temperature Physics; Apr2020, Vol. 199 Issue 1/2, p200-205, 6p
- Publication Year :
- 2020
-
Abstract
- In this paper, the design and test results of a front-end cryogenic amplifier for a bolometer detector are presented. The amplifier is implemented with two low-noise Si JFETs, where the first FET is configured as a source follower amplifier and the second FET provides a high dynamic resistance to keep the voltage gain very close to unity. A 3 dB bandwidth over a wide range from DC to 10 MHz is achieved in this design. The input voltage noise density is ~ 1.9 nV/√Hz at 300 K, which further reduces to 1.6 nV/√Hz at 120 K. The flicker corner frequency observed is below 50 Hz. [ABSTRACT FROM AUTHOR]
- Subjects :
- BOLOMETERS
PREAMPLIFIERS
DETECTORS
LOW noise amplifiers
TEST design
Subjects
Details
- Language :
- English
- ISSN :
- 00222291
- Volume :
- 199
- Issue :
- 1/2
- Database :
- Complementary Index
- Journal :
- Journal of Low Temperature Physics
- Publication Type :
- Academic Journal
- Accession number :
- 142665196
- Full Text :
- https://doi.org/10.1007/s10909-019-02318-z