DVH analysis using a transmission detector and model‐based dose verification system as a comprehensive pretreatment QA tool for VMAT plans: Clinical experience and results.

Authors :
Mohamed Yoosuf, Ahamed B.
AlShehri, Salem
Alhadab, Abdulrahman
Alqathami, Mamdooh
Source :
Journal of Applied Clinical Medical Physics; Nov2019, Vol. 20 Issue 11, p80-87, 8p
Publication Year :
2019

Abstract

Purpose: Dose volume histogram (DVH)‐based analysis is utilized as a pretreatment quality assurance tool to determine clinical relevance from measured dose which is difficult in conventional gamma‐based analysis. In this study, we report our clinical experience with an ionization‐based transmission detector and model‐based verification system, using DVH analysis, as a comprehensive pretreatment QA tool for complex volumetric modulated arc therapy plans. Methods and Materials: Seventy‐three subsequent treatment plans categorized into four clinical sites (Head and Neck, Thorax, Abdomen, and Pelvis) were evaluated. The average dose (Dmean) and dose received by 1% (D1) of the planning target volumes (PTVs) and organs at risks (OARs) calculated using the treatment planning system (TPS) were compared to a computed (model‐based) and reconstructed dose, from the measured fluence, using DVH analysis. The correlation between gamma (3% 3 mm) and DVH‐based analysis for targets was evaluated. Furthermore, confidence and action limits for detector and verification systems were established. Results: Linear regression confirmed an excellent correlation between TPS planned and computed dose using a model‐based verification system (r2 = 1). The average percentage difference between TPS calculated and reconstructed dose for PTVs achieved using DVH analysis for each site is as follows: Head and Neck — 0.57 ± 2.8% (Dmean) and 2.6 ± 2.7% (D1), Abdomen — 0.19 ± 2.8% and 1.64 ± 2.2%, Thorax — 0.24 ± 2.1% and 3.12 ± 2.8%, Pelvis 0.37 ± 2.4% and 1.16 ± 2.3%, respectively. The average percentage of passed gamma values achieved was above 95% for all cases. However, no correlation was observed between gamma passing rates and DVH difference (%) for PTVs (r2 = 0.11). The results demonstrate a confidence limit of 5% (Dmean and D1) for PTVs using DVH analysis for both computed and reconstructed dose distribution. Conclusion: DVH analysis of treatment plan using a model‐based verification system and transmission detector provided useful information on clinical relevance for all cases and could be used as a comprehensive pretreatment patient‐specific QA tool. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
15269914
Volume :
20
Issue :
11
Database :
Complementary Index
Journal :
Journal of Applied Clinical Medical Physics
Publication Type :
Academic Journal
Accession number :
139569891
Full Text :
https://doi.org/10.1002/acm2.12743