Back to Search Start Over

Pipeline Embolization Device for Salvage Treatment of a Willis Covered Stent Prolapse Into the Aneurysmal Sac.

Authors :
Wu, Zeng-Bao
Wang, Sheng
Song, Li-Gang
Yang, Xin-Jian
Mu, Shi-Qing
Source :
Frontiers in Neurology; 10/17/2019, p1-5, 5p
Publication Year :
2019

Abstract

The Willis covered stent (WCS) may prolapse into the aneurysmal sac due to device migration or foreshortening. We present a useful salvage strategy that can reorient a prolapsed WCS into a more suitable alignment. An intra-procedural prolapse of a WCS into a large cavernous aneurysm occurred in a 70-year-old female patient. A pipeline embolization device (PED) was used to retrieve the WCS and successfully accomplish flow diversion. Maintaining proximal access and ensuring that the microwire is securely held within the central axis of the herniated stent are critical until the entire parent vessel can be reconstructed. This salvage technique may help to regain proximal access and reposition the flow diversion constructs following WCS prolapse. [ABSTRACT FROM AUTHOR]

Subjects

Subjects :
WASTE salvage
WOMEN patients

Details

Language :
English
ISSN :
16642295
Database :
Complementary Index
Journal :
Frontiers in Neurology
Publication Type :
Academic Journal
Accession number :
139199651
Full Text :
https://doi.org/10.3389/fneur.2019.01099