Quantitative CT-based bone strength parameters for the prediction of novel spinal implant stability using resonance frequency analysis: a cadaveric study involving experimental micro-CT and clinical multislice CT.

Authors :
Nakashima, Daisuke
Ishii, Ken
Nishiwaki, Yuji
Kawana, Hiromasa
Jinzaki, Masahiro
Matsumoto, Morio
Nakamura, Masaya
Nagura, Takeo
Source :
European Radiology Experimental; 1/22/2019, Vol. 3 Issue 1, pN.PAG-N.PAG, 1p
Publication Year :
2019

Abstract

Background: To predict conventional test forces (peak torque and pull-out force) and a new test force (implant stability quotient [ISQ] value of a spinal pedicle screw) from computed tomography (CT) parameters, including micro-architectural parameters, using high-resolution micro-CT and clinical multislice CT (MSCT) in human cadaveric vertebrae. Methods: Micro-CT scans before/after screw insertion (n = 68) and MSCT scans before screw insertion (n = 58) of human cadaveric vertebrae were assessed for conventional test forces and ISQ value. Three-dimensional volume position adjustment between pre-insertion micro-CT and MSCT scans and post-insertion scans (micro-CT) was performed to extract the volume of the cancellous bone surrounding the pedicle screw. The following volume bone mineral density and micro-architectural parameters were calculated: bone volume fraction, bone surface density (bone surface/total volume (BS/TV)), trabecular thickness, trabecular separation, trabecular number, structure model index, and number of nodes (branch points) of the cancellous bone network/total volume (NNd/TV) using Spearman's rank correlation coefficient with Bonferroni correction. Results: Conventional test forces showed the strongest correlation with BS/TV: peak torque, ρ = 0.811, p = 4.96 × 10<superscript>−17</superscript>(micro-CT) and ρ = 0.730, p = 7.87 × 10<superscript>−11</superscript> (MSCT); pull-out force, ρ = 0.730, p = 1.64 × 10<superscript>−12</superscript> (micro-CT) and ρ = 0.693, p = 1.64 × 10<superscript>−9</superscript> (MSCT). ISQ value showed the strongest correlation with NNd/TV: ρ = 0.607, p = 4.01 × 10<superscript>−8</superscript> (micro-CT) and ρ = 0.515, p = 3.52 × 10<superscript>−5</superscript> (MSCT). Conclusions: Test forces, including the ISQ value, can be predicted using micro-CT and MSCT parameters. This is useful for establishing a preoperative fixation strength evaluation system. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
25099280
Volume :
3
Issue :
1
Database :
Complementary Index
Journal :
European Radiology Experimental
Publication Type :
Academic Journal
Accession number :
137442890
Full Text :
https://doi.org/10.1186/s41747-018-0080-3