Back to Search Start Over

Chiral separation and modeling of quinolones on teicoplanin macrocyclic glycopeptide antibiotics CSP.

Authors :
Ali, Imran
Suhail, Mohd
Asnin, Leonid
Source :
Chirality; Dec2018, Vol. 30 Issue 12, p1304-1311, 8p
Publication Year :
2018

Abstract

New chiral high‐performance liquid chromatography (HPLC) method for the enantiomeric resolution of quinolones is developed and described. The column used was Chirobiotic T (150 × 4.6 mm, 5.0 μm). Three mobile phases used were MeOH:ACN:Water:TEA (70:10:20:0.1%), (60:30:10:0.1%), and (50:30:20:0.1%). The flow rate of the mobile phases was 1.0 mL/min with UV detection at different wavelengths. The values of retention, resolution, and separation factors ranged from 1.5 to 6.0, 1.80 to 2.25, and 2.86 to 6.0, respectively. The limit of detection and quantification ranged from 4.0 to 12 ng and 40 to 52 ng, respectively. The modeling studies indicated strong interactions of R‐enantiomers with teicoplanin chiral selector than S‐enantiomers. The supra molecular mechanism of the chiral recognition was established by modeling and chromatographic studies. It was observed that hydrogen bondings and π‐π interactions are the major forces for chiral separation. The present chiral HPLC method may be used for enantiomeric resolution of quinolones in any matrices. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
08990042
Volume :
30
Issue :
12
Database :
Complementary Index
Journal :
Chirality
Publication Type :
Academic Journal
Accession number :
133218919
Full Text :
https://doi.org/10.1002/chir.23024