Back to Search Start Over

Three-Component Constant Temperature Anemometer CTA3-2: Calibration and Application.

Authors :
Budovsky, A. D.
Sidorenko, A. A.
Polivanov, P. A.
Kornev, E. V.
Salenko, S. D.
Obuhovsky, A. D.
Source :
AIP Conference Proceedings; 2018, Vol. 2027 Issue 1, p030139-1-030139-7, 7p
Publication Year :
2018

Abstract

The study deals with three-component hot-wire anemometer designed and manufactured in ITAM SB RAS. The paper presents the calibration technique of designed and manufactured three-wire probes and describes the data obtained in testing experiments on measurements of steady and unsteady parameters of the flow in subsonic turbulent jet. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
0094243X
Volume :
2027
Issue :
1
Database :
Complementary Index
Journal :
AIP Conference Proceedings
Publication Type :
Conference
Accession number :
132859442
Full Text :
https://doi.org/10.1063/1.5065233