Back to Search Start Over

Crosstalk in an FDM Laboratory Setup and the Athena X-IFU End-to-End Simulator.

Authors :
den Hartog, R.
Kirsch, C.
de Vries, C.
Akamatsu, H.
Dauser, T.
Peille, P.
Cucchetti, E.
Jackson, B.
Bandler, S.
Smith, S.
Wilms, J.
Source :
Journal of Low Temperature Physics; Nov2018, Vol. 193 Issue 3/4, p533-538, 6p
Publication Year :
2018

Abstract

The impact of various crosstalk mechanisms on the performance of the Athena X-IFU instrument has been assessed with detailed end-to-end simulations. For the crosstalk in the electrical circuit, a detailed model has been developed. In this contribution, we test this model against measurements made with an FDM laboratory setup and discuss the assumption of deterministic crosstalk in the context of the weak link effect in the detectors. We conclude that crosstalk levels predicted by the model are conservative with respect to the observed levels. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
00222291
Volume :
193
Issue :
3/4
Database :
Complementary Index
Journal :
Journal of Low Temperature Physics
Publication Type :
Academic Journal
Accession number :
132323706
Full Text :
https://doi.org/10.1007/s10909-018-1901-8