Back to Search Start Over

The Trajectory Simulation and Optimization of Ion Source Chimney for SC200 Cyclotron.

Authors :
Shiwen Xu
Yuntao Song
Liusi Sheng
Gen Chen
Yanfang Bi
Yonghua Chen
Kaizhong Ding
Qingxi Yang
Yahong Xie
Yuqian Chen
Chongdong Hu
Karamyshev, Oleg
Karamysheva, Galina
Morozov, Nikolay
Samsonov, Evgeny Vasilievich
Shirkov, Grigori
Source :
AIP Conference Proceedings; 2018, Vol. 2011 Issue 1, p1-3, 3p
Publication Year :
2018

Abstract

SC200 is an isochronous cyclotron which generate 200 MeV, 500 nA proton for particle therapy. As an important component of the cyclotron, the ion source chimney needs to be tested and optimized. The simulation results and optimization of ion source in test bench for SC200 are described in this paper. The simulation results show that the extraction slit with different sizes and shapes has an influence on the emittance of the extraction beam. To verify the simulation results, the performance of the designed ion source chimney with optimized slits was measured, including the beam current intensity and the beam emittance. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
0094243X
Volume :
2011
Issue :
1
Database :
Complementary Index
Journal :
AIP Conference Proceedings
Publication Type :
Conference
Accession number :
131918760
Full Text :
https://doi.org/10.1063/1.5053391