Online pretreatment verification of high-dose rate brachytherapy using an imaging panel.

Authors :
Gabriel P Fonseca
Mark Podesta
Murillo Bellezzo
Michiel R Van den Bosch
Ludy Lutgens
Ben G L Vanneste
Robert Voncken
Evert J Van Limbergen
Brigitte Reniers
Frank Verhaegen
Source :
Physics in Medicine & Biology; 7/7/2017, Vol. 62 Issue 13, p1-1, 1p
Publication Year :
2017

Abstract

Brachytherapy is employed to treat a wide variety of cancers. However, an accurate treatment verification method is currently not available. This study describes a pre-treatment verification system that uses an imaging panel (IP) to verify important aspects of the treatment plan. A detailed modelling of the IP was only possible with an extensive calibration performed using a robotic arm. Irradiations were performed with a high dose rate (HDR) <superscript>192</superscript>Ir source within a water phantom. An empirical fit was applied to measure the distance between the source and the detector so 3D Cartesian coordinates of the dwell positions can be obtained using a single panel. The IP acquires 7.14 fps to verify the dwell times, dwell positions and air kerma strength (Sk). A gynecological applicator was used to create a treatment plan that was registered with a CT image of the water phantom used during the experiments for verification purposes. Errors (shifts, exchanged connections and wrong dwell times) were simulated to verify the proposed verification system. Cartesian source positions (panel measurement plane) have a standard deviation of about 0.02 cm. The measured distance between the source and the panel (z-coordinate) have a standard deviation up to 0.16 cm and maximum absolute error of  ≈0.6 cm if the signal is close to sensitive limit of the panel. The average response of the panel is very linear with Sk. Therefore, Sk measurements can be performed with relatively small errors. The measured dwell times show a maximum error of 0.2 s which is consistent with the acquisition rate of the panel. All simulated errors were clearly identified by the proposed system. The use of IPs is not common in brachytherapy, however, it provides considerable advantages. It was demonstrated that the IP can accurately measure Sk, dwell times and dwell positions. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
00319155
Volume :
62
Issue :
13
Database :
Complementary Index
Journal :
Physics in Medicine & Biology
Publication Type :
Academic Journal
Accession number :
123568307
Full Text :
https://doi.org/10.1088/1361-6560/aa7028