Back to Search Start Over

A compact permanent-magnet system for measuring magnetic circular dichroism in resonant inelastic soft X-ray scattering.

Authors :
Miyawaki, Jun
Suga, Shigemasa
Fujiwara, Hidenori
Niwa, Hideharu
Kiuchi, Hisao
Harada, Yoshihisa
Source :
Journal of Synchrotron Radiation; Mar2017, Vol. 24 Issue 2, p449-455, 6p
Publication Year :
2017

Abstract

A compact and portable magnet system for measuring magnetic dichroism in resonant inelastic soft X-ray scattering (SX-RIXS) has been developed at the beamline BL07LSU in SPring-8. A magnetic circuit composed of Nd-Fe-B permanent magnets, which realised ∼0.25 T at the center of an 11 mm gap, was rotatable around the axis perpendicular to the X-ray scattering plane. Using the system, a SX-RIXS spectrum was obtained under the application of the magnetic field at an angle parallel, nearly 45° or perpendicular to the incident X-rays. A dedicated sample stage was also designed to be as compact as possible, making it possible to perform SX-RIXS measurements at arbitrary incident angles by rotating the sample stage in the gap between the magnetic poles. This system enables facile studies of magnetic dichroism in SX-RIXS for various experimental geometries of the sample and the magnetic field. A brief demonstration of the application is presented. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
09090495
Volume :
24
Issue :
2
Database :
Complementary Index
Journal :
Journal of Synchrotron Radiation
Publication Type :
Academic Journal
Accession number :
121479564
Full Text :