Back to Search
Start Over
Design of the opacity spectrometer for opacity measurements at the National Ignition Facility.
- Source :
- Review of Scientific Instruments; Nov2016, Vol. 87 Issue 11, p11D623-1-11D623-3, 3p, 2 Diagrams, 4 Graphs
- Publication Year :
- 2016
-
Abstract
- Recent experiments at the Sandia National Laboratory Z facility have called into question models used in calculating opacity, of importance for modeling stellar interiors. An effort is being made to reproduce these results at the National Ignition Facility (NIF). These experiments require a new X-ray opacity spectrometer (OpSpec) spanning 540 eV-2100 eV with a resolving power E/ΔE > 700. The design of the OpSpec is presented. Photometric calculations based on expected opacity data are also presented. First use on NIF is expected in September 2016. [ABSTRACT FROM AUTHOR]
- Subjects :
- OPACITY (Optics)
SPECTROMETERS
X-rays
Subjects
Details
- Language :
- English
- ISSN :
- 00346748
- Volume :
- 87
- Issue :
- 11
- Database :
- Complementary Index
- Journal :
- Review of Scientific Instruments
- Publication Type :
- Academic Journal
- Accession number :
- 119958862
- Full Text :
- https://doi.org/10.1063/1.4962819