Back to Search Start Over

A flow cell for measuring X-ray Compton scattering of liquid at temperatures up to 623 K and pressures up to 20 MPa.

Authors :
Takumi Ono
Masaru Watanabe
Yoshiyuki Sato
Hiroshi Inomata
Koichi Nakahara
Masayoshi Itou
Yoshiharu Sakurai
Source :
Review of Scientific Instruments; Aug2016, Vol. 87 Issue 8, p083106-1-083106-4, 4p, 2 Diagrams, 2 Graphs
Publication Year :
2016

Abstract

A flow-type cell was developed for measuring Compton scattering spectra of heat-sensitive aqueous solution. Compton scattering spectra of water and ethanol were measured in the region from ambient conditions to 623 K and 20 MPa. Compton profiles derived from measurement with the flow-type cell were comparable with those in the literature. Results obtained from the flow-type cell showed that delocalization of electronic charge density of water and ethanol at high temperatures occurred. Delocalization of the electronic charge density of ethanol was greater than that of water at high temperature, which is consistent with the prior works that use proton NMR chemical shifts to describe hydrogen bonding. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
00346748
Volume :
87
Issue :
8
Database :
Complementary Index
Journal :
Review of Scientific Instruments
Publication Type :
Academic Journal
Accession number :
117844777
Full Text :
https://doi.org/10.1063/1.4960706