Back to Search
Start Over
INFLUENCE OF LOCALIZER AND SCAN DIRECTION ON THE DOSE-REDUCING EFFECT OFAUTOMATIC TUBE CURRENT MODULATION IN COMPUTED TOMOGRAPHY.
- Source :
- Radiation Protection Dosimetry; Jun2016, Vol. 169 Issue 1-4, p136-142, 7p
- Publication Year :
- 2016
-
Abstract
- The purpose of this study was to investigate the influence of the localizer and scan direction on the dose-reducing efficacy of the automatic tube current modulation (ATCM) in computed tomography (CT). Craniocaudal and caudocranial chest CT scans, based on anterior-posterior (AP), posterior-anterior (PA), lateral (LAT) or dual AP/LAT localizers, of an anthropomorphic phantom containing thermoluminescent dosimeters (TLDs), were made on three Siemens systems. TLD readings were converted to lung and thyroid doses. A second dose estimation was performed based on Monte Carlo simulations. In addition, the ATCM behaviour of GE and Toshiba was evaluated based on AP, PA and LAT localizers. Compared with AP, tube currents of PA and AP/LAT scans were on average 20 % higher and 40 % lower, respectively, for the Siemens systems. Consequently, thyroid and lung doses increased with 60 % with a PA instead of an AP/LAT scan, with significant differences in image noise. Moreover, the thyroid dose halves by taking the scan in caudocranial direction. Noise values were not significantly different when changing scan direction. [ABSTRACT FROM AUTHOR]
Details
- Language :
- English
- ISSN :
- 01448420
- Volume :
- 169
- Issue :
- 1-4
- Database :
- Complementary Index
- Journal :
- Radiation Protection Dosimetry
- Publication Type :
- Academic Journal
- Accession number :
- 116107765
- Full Text :
- https://doi.org/10.1093/rpd/ncw077