Back to Search Start Over

In vivo real-time assessment of colorectal polyp histology using an optical biopsy forceps system based on laser-induced fluorescence spectroscopy.

Authors :
Rath, Timo
Tontini, Gian E.
Vieth, Michael
Nägel, Andreas
Neurath, Markus F.
Neumann, Helmut
Source :
Endoscopy; 2016, Vol. 48 Issue 6, p557-562, 6p
Publication Year :
2016

Abstract

Background and study aims: In order to reduce time, costs, and risks associated with resection of diminutive colorectal polyps, the American Society for Gastrointestinal Endoscopy (ASGE) recently proposed performance thresholds that new technologies should meet for the accurate real-time assessment of histology of colorectal polyps. In this study, we prospectively assessed whether laser-induced fluorescence spectroscopy (LIFS), using the new WavSTAT4 optical biopsy system, can meet the ASGE criteria. Patients and methods: 27 patients undergoing screening or surveillance colonoscopy were included. The histology of 137 diminutive colorectal polyps was predicted in real time using LIFS and findings were compared with the results of conventional histopathological examination. The accuracy of predicting polyp histology with WavSTAT4 was assessed according to the ASGE criteria. Results: The overall accuracy of LIFS using WavSTAT4 for predicting polyp histology was 84.7 % with sensitivity, specificity, and negative predictive value (NPV) of 81.8 %, 85.2 %, and 96.1 %. When only distal colorectal diminutive polyps were considered, the NPV for excluding adenomatous histology increased to 100 % (accuracy 82.4 %, sensitivity 100 %, specificity 80.6 %). On-site, LIFS correctly predicted the recommended surveillance intervals with an accuracy of 88.9 % (24/27 patients) when compared with histology-based United States guideline recommendations; in the 3 patients for whom LIFS- and histopathology-based recommended surveillance intervals differed, LIFS predicted shorter surveillance intervals. Conclusions: From the data of this pilot study, LIFS using the WavSTAT4 system appears accurate enough to allow distal colorectal polyps to be left in place and nearly reaches the threshold to “resect and discard” them without pathologic assessment. WavSTAT4 therefore has the potential to reduce costs and risks associated with the removal of diminutive colorectal polyps. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
0013726X
Volume :
48
Issue :
6
Database :
Complementary Index
Journal :
Endoscopy
Publication Type :
Academic Journal
Accession number :
115818307
Full Text :
https://doi.org/10.1055/s-0042-102251