Fixed-wavelength H2O absorption spectroscopy system enhanced by an on-board external-cavity diode laser.

Authors :
Mack S Brittelle
Jean M Simms
Scott T Sanders
James R Gord
Sukesh Roy
Source :
Measurement Science & Technology; Mar2016, Vol. 27 Issue 3, p1-1, 1p
Publication Year :
2016

Abstract

We describe a system designed to perform fixed-wavelength absorption spectroscopy of H<subscript>2</subscript>O vapor in practical combustion devices. The system includes seven wavelength-stabilized distributed feedback (WSDFB) lasers, each with a spectral accuracy of  ±1 MHz. An on-board external cavity diode laser (ECDL) that tunes 1320–1365 nm extends the capabilities of the system. Five system operation modes are described. In one mode, a sweep of the ECDL is used to monitor each WSDFB laser wavelength with an accuracy of  ±30 MHz. Demonstrations of fixed-wavelength thermometry at 10 kHz bandwidth in near-room-temperature gases are presented; one test reveals a temperature measurement error of ~0.43%. [ABSTRACT FROM AUTHOR]