Back to Search Start Over

Development of a handheld fluorescence imaging camera for intraoperative sentinel lymph node mapping.

Authors :
Szyc, Łukasz
Bonifer, Stefanie
Walter, Alfred
Jagemann, Uwe
Grosenick, Dirk
Macdonald, Rainer
Source :
Journal of Biomedical Optics; May2015, Vol. 20 Issue 5, p1-10, 10p
Publication Year :
2015

Abstract

We present a compact fluorescence imaging system developed for real-time sentinel lymph node mapping. The device uses two near-infrared wavelengths to record fluorescence and anatomical images with a single charge-coupled device camera. Experiments on lymph node and tissue phantoms confirmed that the amount of dye in superficial lymph nodes can be better estimated due to the absorption correction procedure integrated in our device. Because of the camera head's small size and low weight, all accessible regions of tissue can be reached without the need for any adjustments. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
10833668
Volume :
20
Issue :
5
Database :
Complementary Index
Journal :
Journal of Biomedical Optics
Publication Type :
Academic Journal
Accession number :
108988200
Full Text :
https://doi.org/10.1117/1.JBO.20.5.051025