Back to Search Start Over

Your lab focus. A comparison of blood collection in glass versus plastic vacutainers on results of esoteric coagulation assays.

Authors :
Flanders MM
Crist R
Rodgers GM
Source :
Laboratory Medicine; Oct2003, Vol. 34 Issue 10, p732-735, 4p
Publication Year :
2003

Abstract

We compared the BD standard glass sodium citrate Vacutainer to the BD plastic sodium citrate Vacutainer, evaluating 19 esoteric coagulation analytes, including comparisons, mean, and standard deviation for 20 normal donor plasma samples.Only 1 analyte, the thrombin time, was found to vary depending on the collection tube used.Laboratories should establish separate reference intervals for plastic and glass Vacutainers for the thrombin time assay. Other esoteric coagulation analytes do not need separate reference intervals. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
00075027
Volume :
34
Issue :
10
Database :
Complementary Index
Journal :
Laboratory Medicine
Publication Type :
Academic Journal
Accession number :
106690784
Full Text :
https://doi.org/10.1309/bmqhcd6b8984rla3